Cavitation from radial shock wave devices versus vibrating massagers · Lab study · 2015

Do radial shockwave devices create collapsing bubbles, or just vibrate like massagers?

In water-tank tests and tiny worms, 2 radial shockwave devices acted differently from a vibrating massager. Both formed collapsing bubbles (cavitation) that pitted X-ray film and paralysed 77% of worms in water, while the massager made none. Faster pulsing (15 vs 1 Hz) made fewer bubbles, and bar varied by brand.

Explores how shockwave works
radial shockwave devices tested against 1 vibrating massager
2
of tiny worms paralysed by radial pulses in water
77%
pulses per second made fewer bubbles than 1 per second
15

Good to know: The senior author is a paid consultant to Electro Medical Systems (maker of one tested device), whose laboratory ran the water-tank measurements.

Császár NB, Angstman NB, Milz S, et al. PLoS One, 2015 · Read the paper ↗

The treatment

Acoustic hydrophone setup (water tank)
Laser fiber optic probe hydrophone (FOPH 2000) in a 300-L demineralized water tank (~5 µS/cm) at Electro Medical Systems' laboratory (Nyon, Switzerland) per IEC-61846:1998. Tested Storz Medical D-Actor 200 (Model 2007, 15-mm applicator) at 3 bar and maximum 5 bar, and EMS Swiss DolorClast ('Radial' handpiece, 15-mm applicator) at 3 bar and maximum 4 bar, at 1, 5, 10, and 20 mm along the applicator axis (5 repetitions averaged). Vibracare (G5/General Physiotherapy, SKU VC24B) could not be measured due to oscillating head motion.
High-speed cavitation imaging and X-ray film tests
At EPFL Lausanne, submerged 15-mm tips of D-Actor 200 (3 and 5 bar, 1 and 15 Hz), Swiss DolorClast (3 and 4 bar, 1 and 15 Hz), and Vibracare (50 cycles/s continuous) were filmed in vacuum-degassed deionized water at 300,000 frames/s (370 ns exposure, triplicates; 1,001 frames = 3.3 ms per single radial pulse; 10,000 frames for Vibracare). At LMU Munich, submerged X-ray sheets 3 mm below each tip received 10,000 pulses at 1 Hz at maximum settings (5 bar D-Actor 200; 4 bar Swiss DolorClast) or 1 hour of Vibracare at 50 cycles/s.
Caenorhabditis elegans locomotion bioassay
Wells containing ~20 adult worms in 300 µL S-Medium (cavitation-permissive) or viscous polyvinyl alcohol (PVA, cavitation-suppressing; n = 5 wells each) received either 500 impulses from the Swiss DolorClast 6-mm applicator sealed with an O-ring at 2 bar (stated by the authors to correspond to 0.08 mJ/mm² in this setup) and 1 Hz, 500 vibrations (10 s at 50 cycles/s) from Vibracare taped beneath sealed plates, or switched-off sham handling. D-Actor 200 was not tested on worms because no 6-mm applicator was available.

Who took part

Water-tank pressure sensors, high-speed cameras, submerged X-ray film, and 99 to 130 microscopic roundworms (Caenorhabditis elegans) per group in liquid or thick gel.

Side effects

The study does not report side effects.

All the numbers and the full appraisal

FROM THE FULL PAPER

What the full paper reports.

No human participants or mammalian tissues. Bench experiments used a 300-L demineralized water tank (IEC-61846:1998), a degassed deionized-water polycarbonate vessel, submerged Agfa STRUCTURIX D4DW X-ray film sheets, and adult wild-type Caenorhabditis elegans nematodes (N2, Bristol; 5 wells of ~20 worms per condition in 96-well plates, yielding 99–130 worms analysed per group in S-Medium or viscous 31,000 g/mol polyvinyl alcohol).

Laser fiber optic probe hydrophone (FOPH 2000, IEC-61846:1998)1 reported outcome
01
~15 µs pulse waveforms averaged over 5 shots at 1, 5, 10, and 20 mm

Acoustic pressure and energy flux density in water (Table 1)

Laser fiber optic probe hydrophone (FOPH 2000, IEC-61846:1998) MPa, µs, and mJ/mm² across bar settings

Both radial devices produced ~15 µs biphasic/oscillating waves with non-linear relationships between input air pressure (bar) and output positive energy flux density (EFD+). At 3 bar and 1 mm distance, D-Actor 200 delivered P+ 8.6 MPa, P− −6.7 MPa, rise time 2.6 µs, and EFD+ 0.10 mJ/mm² (falling to 0.04, 0.02, and <0.01 mJ/mm² at 5, 10, and 20 mm), while Swiss DolorClast delivered P+ 10.1 MPa, P− −5.7 MPa, rise time 2.9 µs, and the same EFD+ 0.10 mJ/mm² (0.04, 0.02, and 0.02 mJ/mm² at 5, 10, and 20 mm). At maximum air pressure (5 bar on D-Actor 200 vs 4 bar on Swiss DolorClast), 1-mm EFD+ was identical at 0.14 mJ/mm² (D-Actor 200: P+ 10.5 MPa, P− −9.0 MPa, Rt 3.0 µs; Swiss DolorClast: P+ 11.3 MPa, P− −6.0 MPa, Rt 3.4 µs; 20-mm EFD+ 0.02 vs 0.01 mJ/mm²). Vibracare could not be measured.

Analysis details and source

Table 1 reports means of 5 shots without SDs. Note axis-unit error in Figure 3 (p. 8): the y-axis is labelled 'Bar' (−10 to 20) even though peak heights (~10) correspond to MPa in Table 1 (1 MPa = 10 bar).

High-speed CCD shadowgraphy (Photron Ultima APX, 300,000 fps)1 reported outcome
02
3.3 ms window around single pulse (radial) or 10,000 frames (Vibracare)

Cavitation bubble formation across pressure and pulse frequency

High-speed CCD shadowgraphy (Photron Ultima APX, 300,000 fps) Bubble timing and binarized pixel counts

Both D-Actor 200 and Swiss DolorClast generated cavitation bubbles starting 10 µs after the wave front, peaking around 120 µs later, and lasting ~1 ms. Cavitation increased with higher bar settings, but non-linearly: at 3 bar D-Actor 200 generated substantially more cavitation pixels than Swiss DolorClast, whereas at maximum settings (5 bar vs 4 bar) peak pixel counts were comparable (Figure 6). Both devices produced larger bubbles and more cavitation pixels at 1 Hz than at 15 Hz (drop more pronounced on D-Actor 200). Vibracare produced zero cavitation bubbles (Figure 7).

Analysis details and source

2D maximum-intensity shadowgraphy projections could not separate overlapping bubbles along the optical axis or count small versus large bubbles individually, preventing formal statistical testing between the two radial devices.

Light microscopy of exposed Agfa STRUCTURIX D4DW X-ray film1 reported outcome
03
After 10,000 radial pulses at 1 Hz or 1 hour of Vibracare at 50 cycles/s

Mechanical surface damage on submerged X-ray film

Light microscopy of exposed Agfa STRUCTURIX D4DW X-ray film Qualitative surface pitting and silver-grain blackening

At 3 mm distance in water, 10,000 pulses at maximum settings produced a central impression surrounded by a blackened ring for both radial devices: D-Actor 200 (5 bar) completely perforated a central hole through the film, while Swiss DolorClast (4 bar) indented the film without complete perforation. One hour of Vibracare (50 cycles/s) left no detectable mark (Figure 8).

Analysis details and source

Qualitative microscopy; the number of replicate X-ray sheets tested per device is not stated.

WormLab 1-minute video tracking on agar plates1 reported outcome
04
Immediately after 500 impulses (Swiss DolorClast) or 500 vibrations (Vibracare)

Caenorhabditis elegans locomotion speed and paralysis in S-Medium vs PVA

WormLab 1-minute video tracking on agar plates Mean ± SEM locomotion speed (µm/s) and % paralyzed worms

In cavitation-permissive S-Medium, 500 impulses from Swiss DolorClast (6-mm tip, 2 bar, 1 Hz) reduced mean worm locomotion speed to 8 ± 1 µm/s (n = 100 worms; 77% paralyzed) compared with sham controls at 88 ± 6 µm/s (n = 130; 18% paralyzed; Dunn's test p < 0.001) and Vibracare at 84 ± 6 µm/s (n = 110; 8% paralyzed; p < 0.001 vs radial shock waves). In cavitation-suppressing viscous PVA, speeds did not differ significantly across sham control (57 ± 5 µm/s, n = 99; 17% paralyzed), Swiss DolorClast (47 ± 4 µm/s, n = 104; 29% paralyzed), and Vibracare (59 ± 5 µm/s, n = 100; 17% paralyzed), attributing the acute worm impairment primarily to cavitation.

Analysis details and source

Non-human nematode assay; spreads are standard errors of the mean (SEM), not SDs, even though Kruskal–Wallis and Dunn's non-parametric tests were used after normality failed. Individual worms (99–130) pooled across 5 wells per group served as the unit of analysis, and D-Actor 200 was not tested in this assay.

How treatment was delivered
Acoustic hydrophone setup (water tank)

Laser fiber optic probe hydrophone (FOPH 2000) in a 300-L demineralized water tank (~5 µS/cm) at Electro Medical Systems' laboratory (Nyon, Switzerland) per IEC-61846:1998. Tested Storz Medical D-Actor 200 (Model 2007, 15-mm applicator) at 3 bar and maximum 5 bar, and EMS Swiss DolorClast ('Radial' handpiece, 15-mm applicator) at 3 bar and maximum 4 bar, at 1, 5, 10, and 20 mm along the applicator axis (5 repetitions averaged). Vibracare (G5/General Physiotherapy, SKU VC24B) could not be measured due to oscillating head motion.

High-speed cavitation imaging and X-ray film tests

At EPFL Lausanne, submerged 15-mm tips of D-Actor 200 (3 and 5 bar, 1 and 15 Hz), Swiss DolorClast (3 and 4 bar, 1 and 15 Hz), and Vibracare (50 cycles/s continuous) were filmed in vacuum-degassed deionized water at 300,000 frames/s (370 ns exposure, triplicates; 1,001 frames = 3.3 ms per single radial pulse; 10,000 frames for Vibracare). At LMU Munich, submerged X-ray sheets 3 mm below each tip received 10,000 pulses at 1 Hz at maximum settings (5 bar D-Actor 200; 4 bar Swiss DolorClast) or 1 hour of Vibracare at 50 cycles/s.

Caenorhabditis elegans locomotion bioassay

Wells containing ~20 adult worms in 300 µL S-Medium (cavitation-permissive) or viscous polyvinyl alcohol (PVA, cavitation-suppressing; n = 5 wells each) received either 500 impulses from the Swiss DolorClast 6-mm applicator sealed with an O-ring at 2 bar (stated by the authors to correspond to 0.08 mJ/mm² in this setup) and 1 Hz, 500 vibrations (10 s at 50 cycles/s) from Vibracare taped beneath sealed plates, or switched-off sham handling. D-Actor 200 was not tested on worms because no 6-mm applicator was available.

HARMS & TOLERABILITY

No human participants were exposed, so no patient harms were collected. Based on cavitation generation in water and nematode paralysis, the authors emphasize that radial shock wave devices can cause hematomas, vessel rupture, or organ injury and are contraindicated over air-filled tissues such as the lungs (unlike respiratory percussors such as Vibracare). Competing interests and laboratory setting: senior author Christoph Schmitz serves as a paid consultant for and receives benefits from Electro Medical Systems (manufacturer/distributor of Swiss DolorClast), and the hydrophone measurements were carried out in Electro Medical Systems' laboratory in Nyon; co-author John P. Furia was salaried for clinical work at SUN Orthopaedic Group.

LIMITS OF THE ANALYSIS

Author-reported: Vibracare pressure waveforms could not be measured with the laser hydrophone because its head oscillates; 2D high-speed shadowgraphy represents maximum-intensity projections that mask overlapping bubbles along the camera axis and prevent separate counting of small and large bubbles, precluding an in-depth quantitative cavitation comparison between the two radial devices; and the physical reason why 15 Hz produced smaller cavitation fields than 1 Hz remains unknown.

Extractor interpretation: device-maker consultancy and facility use — the senior author is a paid consultant to Electro Medical Systems (Swiss DolorClast), whose laboratory hosted the hydrophone comparison against competitor Storz Medical's D-Actor 200; D-Actor 200 was omitted from the C. elegans assay for lack of a 6-mm tip; worm exposure geometries differed between direct submerged O-ring contact (DolorClast) and external plate vibration through parafilm (Vibracare), with individual worms pooled across 5 wells; Figure 3 mislabels MPa values as 'Bar' on its y-axis; and free-water cavitation and nematode paralysis do not establish cavitation thresholds or clinical efficacy in human tendons or muscles.

Online Supporting Information (S1 File, S1 Fig, S2 Fig) covering BTL-5000 SWT Power, Swiss DolorClast EvoBlue and Power+ handpieces, and Zimmer en Puls V. 2.0 was not part of the local PDF and was not read.

Findings checked against the full source by a separate automated reviewer.

Limitations noted by the reviewers

  • Author-reported: Vibracare pressure waveforms could not be measured with the laser hydrophone because its head oscillates; 2D high-speed shadowgraphy represents maximum-intensity projections that mask overlapping bubbles along the camera axis and prevent separate counting of small and large bubbles, precluding an in-depth quantitative cavitation comparison between the two radial devices; and the physical reason why 15 Hz produced smaller cavitation fields than 1 Hz remains unknown.
  • Extractor interpretation: device-maker consultancy and facility use — the senior author is a paid consultant to Electro Medical Systems (Swiss DolorClast), whose laboratory hosted the hydrophone comparison against competitor Storz Medical's D-Actor 200; D-Actor 200 was omitted from the C. elegans assay for lack of a 6-mm tip; worm exposure geometries differed between direct submerged O-ring contact (DolorClast) and external plate vibration through parafilm (Vibracare), with individual worms pooled across 5 wells; Figure 3 mislabels MPa values as 'Bar' on its y-axis; and free-water cavitation and nematode paralysis do not establish cavitation thresholds or clinical efficacy in human tendons or muscles.
  • Online Supporting Information (S1 File, S1 Fig, S2 Fig) covering BTL-5000 SWT Power, Swiss DolorClast EvoBlue and Power+ handpieces, and Zimmer en Puls V. 2.0 was not part of the local PDF and was not read.
Research review

Source-linked critical reading

Laboratory physics and nematode bench study

Review statusAI-assisted draft · independent automated checkThis automated check comes before the editor’s clinical review.
Study design

Bench physics and invertebrate laboratory study combining IEC-61846:1998 laser fiber optic probe hydrophone (FOPH) measurements in demineralized water, 300,000-fps shadowgraphic cavitation imaging in degassed deionized water, mechanical pitting of submerged X-ray film, and a Caenorhabditis elegans locomotion assay in cavitation-permissive (S-Medium) versus cavitation-suppressing (polyvinyl alcohol, PVA) liquid. No human participants or mammalian tissues were studied.

Participants and setting

No human participants. Experiments were conducted across three sites: water-tank hydrophone tests at Electro Medical Systems (Nyon, Switzerland), high-speed camera imaging at EPFL (Lausanne, Switzerland), and X-ray film and wild-type C. elegans (N2, Bristol; 5 wells of ~20 worms per group, 99–130 worms tracked per condition) assays at Ludwig-Maximilians-University of Munich.

Treatment protocol

Two air-driven ballistic radial devices with 15-mm applicators — Storz Medical D-Actor 200 (operated at 3 bar and maximum 5 bar; 1 and 15 Hz) and EMS Swiss DolorClast ('Radial' handpiece at 3 bar and maximum 4 bar; 1 and 15 Hz) — tested at 1–20 mm for hydrophone waveforms, single pulses for high-speed cavitation imaging, and 10,000 pulses at 1 Hz (3 mm above X-ray film). In C. elegans, only the Swiss DolorClast 6-mm applicator was tested (500 pulses at 2 bar, stated as 0.08 mJ/mm² for that tip, at 1 Hz).

Comparison

Vibracare rotating-flywheel percussive/vibrating massager (G5/General Physiotherapy, operated at maximum 50 cycles/s continuously for imaging, 1 hour over X-ray film, and 10 s = 500 vibrations beneath sealed 96-well worm plates), plus switched-off sham controls and viscous PVA medium in the worm experiments.

What was measured

Peak positive pressure (P+), peak negative pressure (P−), rise time (Rt), and positive energy flux density (EFD+) at 1, 5, 10, and 20 mm; binarized cavitation-bubble pixel counts over 3.3 ms; microscopic X-ray film indentation and silver-grain blackening; and 1-minute agar-plate C. elegans locomotion speed (mean ± SEM) and paralysis rate.

Results and interpretation

Both radial devices generated ~15 µs waves with matching 1-mm EFD+ at 3 bar (0.10 mJ/mm²; P+ 8.6 vs 10.1 MPa; P− −6.7 vs −5.7 MPa) and at maximum air pressure despite different bar ratings (0.14 mJ/mm² at 5 bar on D-Actor 200 vs 4 bar on Swiss DolorClast; P+ 10.5 vs 11.3 MPa; P− −9.0 vs −6.0 MPa). Both formed cavitation bubbles within 10 µs (peaking at ~120 µs, lasting ~1 ms, larger and more extensive at 1 Hz than 15 Hz, and higher for D-Actor 200 at 3 bar but similar at maximum settings) and damaged X-ray film (perforated by D-Actor 200). Vibracare produced neither cavitation nor X-ray film damage. In S-Medium, Swiss DolorClast lowered worm speed to 8 ± 1 µm/s (77% paralyzed) versus 88 ± 6 µm/s in sham controls (18% paralyzed, p < 0.001) and 84 ± 6 µm/s with Vibracare (8% paralyzed), whereas PVA abolished significant speed differences (57 ± 5, 47 ± 4, and 59 ± 5 µm/s).

Harms and disclosures

No human harms were studied; the authors caution that cavitation-generating radial devices must not be applied over air-filled organs such as lungs. Senior author Christoph Schmitz is a paid consultant receiving benefits from Electro Medical Systems (maker of Swiss DolorClast), and the hydrophone tests were performed in Electro Medical Systems' laboratory; co-author John P. Furia was salaried for clinical work at SUN Orthopaedic Group.

Limitations

Hydrophone measurements were performed at one manufacturer's facility by a team whose senior author consults for that manufacturer; Vibracare could not be measured by hydrophone; 2D shadowgraphy precluded 3D bubble counting across radial devices; D-Actor 200 lacked a 6-mm applicator for the worm assay; worm delivery differed (submerged tip vs external plate vibration) and pooled individual worms across 5 wells; Figure 3 mislabels MPa values as 'Bar'; online supplements on three other devices/handpieces were unread; and water/nematode findings do not measure cavitation or clinical effects in human soft tissue.

Applicability

Supports distinguishing ballistic radial shock wave devices from mechanical vibrating massagers for safety contraindications (such as avoiding lung fields) and shows why compressed-air bar settings cannot be equated across brands or converted linearly to mJ/mm². It does not establish clinical dosing or comparative effectiveness in patients.

What stronger evidence needs

Independent acoustic and cavitation measurements in tissue-mimicking media and ex vivo musculoskeletal tissues across devices, handpieces, and frequencies (1 Hz vs 15 Hz), paired with blinded randomized clinical trials testing whether frequency or device differences affect patient pain and function.

THE ORIGINAL SOURCE

Read the original article.

Radial Shock Wave Devices Generate Cavitation.

Császár NB, Angstman NB, Milz S, Sprecher CM, Kobel P, Farhat M, Furia JP, Schmitz C. · PLoS One · 2015

Open the publication record or available full-text files.

Checked against the full paper · Clinical review: Dr. Marcus Yu Bin Pai, · How we review →