Shockwave therapy in orthopaedic disorders: a 2012 narrative overview · Lab study · 2012
How did early lab and clinical studies describe shockwave for bone and tendon disorders?
This 2012 review summarizes rat, rabbit, dog and cell studies linking focused shockwave to blood-vessel and bone-growth signals. In clinic, it mixes single-group series with blinded trials, quoting wide ranges (34%–88% in heel pain; 50%–85% bone union) alongside trials where shockwave matched placebo.
Explores how shockwave works- references across physics, animal and clinical reports
- 129
- focused generator types described
- 3
- year published
- 2012
Good to know: The author served on the scientific advisory committee of Sanuwave, maker of one of the focused shockwave devices discussed.
Wang CJ J Orthop Surg Res, 2012 · Read the paper ↗
The treatment
- Shockwave generation and physical parameters described
- Describes three focused generation principles: electrohydraulic (high-voltage underwater spark discharge with elliptical reflector, large axial focal volume), electromagnetic (coil magnetic field with acoustic focus lens), and piezoelectric (>1,000 spherical piezocrystals self-focusing to a small high-energy focal volume). Contrasts uniphasic shockwaves (peak pressure as high as 500 bar) with biphasic ultrasound (0.5 bar peak). Radial/ballistic devices are omitted from the physics section even though radial clinical studies are cited later.
- Examples of clinical and animal doses cited in text
- Cites Rompe et al.'s rabbit Achilles study warning not to exceed 0.28 mJ/mm² in clinical tendon treatment after 0.60 mJ/mm² caused stronger tissue reaction and potential tendon damage; cites 3 weekly sessions of 1,000 impulses at 0.06 mJ/mm² in plantar fasciitis, 3,000 versus 30 impulses in lateral epicondylitis, 2 × 2,000 or 1,500 impulses in calcific shoulder tendinitis, and 2,000 or 3,000 impulses in bone pseudarthroses. On p. 3, Buchbinder et al.'s 2002 plantar-fasciitis regimen is quoted using that trial's cumulative EFD × impulses convention ('a total dose of at least 1,000 mJ/mm²' versus '6.0 mJ/mm²' in placebo) rather than per-pulse EFD. Also states from unpublished personal observations that local anaesthesia did not alter plantar-fasciitis results and mentions constant sedation if pain is intolerable.
Who took part
No new patients. Draws on 129 earlier reports covering bone marrow cells, rats, rabbits and dogs, plus clinical series and trials in heel, elbow, shoulder, knee, Achilles and bone disorders.
Side effects
Mentions brief skin redness, pinpoint bleeding, mild bruising and migraine without counts; high doses (0.60 mJ/mm²) damaged rabbit Achilles tendons in lab tests.
All the numbers and the full appraisal
What the full paper reports.
No new patient cohort or systematic study sample. Single-author narrative review citing 129 references spanning acoustic generation physics, in vitro marrow stromal cells, rat/rabbit/dog bone and Achilles tendon-bone junction experiments, uncontrolled clinical case series, comparative cohorts, and randomized trials.
Narrative review of preclinical bone, tendon, and marrow-cell studies1 reported outcome
Animal and cell signalling findings in bone, tendon–bone junction, and necrotic femoral head
Narrative review of preclinical bone, tendon, and marrow-cell studies Qualitative preclinical summary
Summarizes rat, dog, and rabbit fracture/non-union studies reporting enhanced callus, cortical bone, bone mineral density, and femur strength at higher shockwave energy (alongside Forriol et al.'s contrasting report of delayed bone healing), linked in marrow stromal cells to superoxide-mediated ERK/CBFA-1 activation, membrane hyperpolarization/Ras, TGF-β1, and VEGF. At the rabbit Achilles tendon–bone junction, eNOS and VEGF rose from week 1 to week 8 before returning to baseline at week 12, while PCNA and neovessels persisted ≥12 weeks; in rabbit femoral-head necrosis, BMP-2 and VEGF rose in subchondral bone.
Analysis details and source
Narrative summary largely featuring the author's own preclinical papers, without sample sizes, quantitative effect estimates, or methodological quality grading.
Narrative literature overview2 reported outcomes
Narrative summary of tendinopathy and plantar-fasciitis studies
Narrative literature overview Reported success-rate ranges and selected study summaries
Reports broad 'success rate' ranges across mixed study designs: 34%–88% in proximal plantar fasciitis (citing positive series including Wang's uncontrolled 79-patient/85-heel cohort alongside Buchbinder 2002, Haake 2003, and Speed 2003 randomized trials where active ESWT performed similarly to placebo), 68%–91% in lateral epicondylitis (citing Rompe 1996 and Wang's 57-patient vs 6-control cohort alongside a 9-trial Cochrane review, a 20-study review, and Speed 2002 finding outcomes similar to sham and short-term advantage for steroid injection), 78%–91% in calcific shoulder tendinitis (citing calcium resorption of 57.6%–62.5% and Wang's 37-patient vs 6-control cohort), and 73.5%–87.5% in patellar and Achilles tendinopathies.
Analysis details and source
Contradiction kept as printed: the Abstract (p. 1) states tendinopathy success rates ranged from 65% to 91%, whereas p. 3 reports 34% to 88% for plantar fasciitis; paired limbs (85 heels in 79 patients, 58 elbows in 57, 39 shoulders in 37, 30 knees in 27) and tiny 6-patient control groups are cited alongside randomized trials without weighting by study design.
Narrative summary of bone non-unions, femoral-head osteonecrosis, and emerging uses
Narrative literature overview Reported union and arthroplasty proportions across cited studies
Cites 50%–85% bony union across mostly single-arm non-union and delayed-union series (including Schaden 85% of 115, Valchanou 70/82, Vogel 60.4% of 48, Wang 82.4% of 72 at 6 months, Rompe 50%, and Elster 80.2% of 172 tibial non-unions). For early femoral-head osteonecrosis, highlights Wang's comparison of ESWT (23 patients, 29 hips) versus core decompression plus non-vascularized fibular grafting (25 patients, 28 hips), reporting total hip arthroplasty in 3% vs 21% at 1 year (P = 0.039), 10% vs 32% at 2 years (P = 0.044), and 24% vs 64% at 8–9 years (P = 0.002), plus a cohort comparing SLE (15 patients, 26 hips) with non-SLE hips (24 patients, 29 hips; arthroplasty 12% vs 14%). Briefly mentions Peyronie's disease, complex regional pain syndrome, rat knee OA, rabbit spinal fusion, tumour cells, gene transfer, diabetic ulcers, and ischaemic heart disease.
Analysis details and source
Paired hips per patient are not adjusted in the cited osteonecrosis percentages, non-union series lack untreated or randomized surgical controls in nearly all cited reports, and the Conclusion extrapolates that ESWT has the potential to replace surgery in many orthopaedic disorders.
How treatment was delivered
Describes three focused generation principles: electrohydraulic (high-voltage underwater spark discharge with elliptical reflector, large axial focal volume), electromagnetic (coil magnetic field with acoustic focus lens), and piezoelectric (>1,000 spherical piezocrystals self-focusing to a small high-energy focal volume). Contrasts uniphasic shockwaves (peak pressure as high as 500 bar) with biphasic ultrasound (0.5 bar peak). Radial/ballistic devices are omitted from the physics section even though radial clinical studies are cited later.
Cites Rompe et al.'s rabbit Achilles study warning not to exceed 0.28 mJ/mm² in clinical tendon treatment after 0.60 mJ/mm² caused stronger tissue reaction and potential tendon damage; cites 3 weekly sessions of 1,000 impulses at 0.06 mJ/mm² in plantar fasciitis, 3,000 versus 30 impulses in lateral epicondylitis, 2 × 2,000 or 1,500 impulses in calcific shoulder tendinitis, and 2,000 or 3,000 impulses in bone pseudarthroses. On p. 3, Buchbinder et al.'s 2002 plantar-fasciitis regimen is quoted using that trial's cumulative EFD × impulses convention ('a total dose of at least 1,000 mJ/mm²' versus '6.0 mJ/mm²' in placebo) rather than per-pulse EFD. Also states from unpublished personal observations that local anaesthesia did not alter plantar-fasciitis results and mentions constant sedation if pain is intolerable.
No adverse-event rates or systematic harms data are reported across conditions. Under proximal plantar fasciitis (p. 4), the review lists local reddening, ecchymosis, mild hematoma, and migraine as self-limiting complications, notes cavitation and high energy (0.60 mJ/mm² in rabbit Achilles tendon) can damage tissue (pp. 2–3), and repeatedly describes clinical complication rates as 'low and negligible' (pp. 1, 4, 5) without numerators or denominators. Competing interests (p. 6): the author served on the scientific advisory committee of Sanuwave (Alpharetta, GA, maker of the OssaTron device highlighted on p. 1), while reporting no honoraria or commercial benefit for writing the manuscript.
Author-reported: the exact biological mechanism of shockwave therapy remains unknown, and clinical controversy persists in plantar fasciitis and lateral epicondylitis across studies using different patient selection criteria, devices, energy levels, local anaesthesia, and outcome measures.
Extractor interpretation: single-author unsystematic narrative review without search methods, selection criteria, risk-of-bias assessment, tables, or confidence intervals; advisory-board tie to Sanuwave; internal mismatch between the Abstract's 65%–91% success claim and lower ranges in the body text (34%–88% in plantar fasciitis, 50%–85% in non-unions); omission of radial devices from the physics section despite citing radial trials; heavy reliance on the author's uncontrolled or 6-control cohorts and unpublished anaesthesia observations; and sweeping conclusions about safety and replacing surgery.
Two published comments on this review indexed in PubMed (PMIDs 23895659 and 23991453, J Orthop Surg Res 2013;8:22) were not included in the local file and were not read.
Findings checked against the full source by a separate automated reviewer.
Limitations noted by the reviewers
- Author-reported: the exact biological mechanism of shockwave therapy remains unknown, and clinical controversy persists in plantar fasciitis and lateral epicondylitis across studies using different patient selection criteria, devices, energy levels, local anaesthesia, and outcome measures.
- Extractor interpretation: single-author unsystematic narrative review without search methods, selection criteria, risk-of-bias assessment, tables, or confidence intervals; advisory-board tie to Sanuwave; internal mismatch between the Abstract's 65%–91% success claim and lower ranges in the body text (34%–88% in plantar fasciitis, 50%–85% in non-unions); omission of radial devices from the physics section despite citing radial trials; heavy reliance on the author's uncontrolled or 6-control cohorts and unpublished anaesthesia observations; and sweeping conclusions about safety and replacing surgery.
- Two published comments on this review indexed in PubMed (PMIDs 23895659 and 23991453, J Orthop Surg Res 2013;8:22) were not included in the local file and were not read.
Source-linked critical reading
Single-author narrative review of animal and clinical studies
This 2012 overview summarizes early laboratory ideas on how shockwaves stimulate blood-vessel and bone-growth signals in animals, and lists orthopaedic uses from heel pain and calcific shoulder tendinitis to fracture non-unions and hip osteonecrosis. However, it mixes uncontrolled case series with blinded placebo trials, omits radial devices from its physics section, quotes different success-rate ranges in the abstract (65%–91%) and body text (down to 34% in plantar fasciitis and 50% in non-unions), and comes from an author on a shockwave manufacturer's advisory board — so condition-specific randomized trials are needed to judge actual patient benefit.
How did a 2012 single-author orthopaedic review summarize shockwave generation principles, animal bone and tendon-junction mechanisms, and clinical results across musculoskeletal disorders?
Single-author unsystematic narrative review citing 129 references across shockwave physics, cell and animal models, case series, comparative cohorts, and randomized trials, without database search criteria, study-selection rules, risk-of-bias evaluation, tables, figures, or meta-analysis.
No new participant cohort or pooled sample total. Reviews rat, rabbit, and dog bone and tendon models alongside published clinical cohorts and trials in plantar fasciitis, lateral epicondylitis, calcific shoulder tendinitis, patellar and Achilles tendinopathy, long-bone non-unions, femoral-head osteonecrosis, and brief non-musculoskeletal uses.
Describes electrohydraulic, electromagnetic, and piezoelectric focused shockwave generators (uniphasic peak pressures up to 500 bar) while omitting radial/ballistic mechanics despite citing radial trials. Mentions diverse regimens from 3 × 1,000 low-energy pulses at 0.06 mJ/mm² in plantar fasciitis to 1,500–3,000 pulses in elbow, shoulder, and bone non-union studies, and cites rabbit Achilles tendon damage risk at 0.60 mJ/mm² (advising ≤0.28 mJ/mm² for tendons).
No systematic comparator. Contrasts shockwaves with biphasic ultrasound (0.5 bar peak) on physical grounds, and narratively cites studies comparing ESWT against sham/placebo, eccentric exercise, physiotherapy, corticosteroid injection, alendronate, plantar fasciotomy, patellar tendon surgery, or core decompression with fibular grafting.
No unified outcome set or effect-size extraction: summarizes animal callus/bone strength and angiogenic/osteogenic markers (eNOS, VEGF, PCNA, BMP-2, CBFA-1, TGF-β1), broad clinical 'success rate' ranges, radiographic calcium disappearance, bony union proportions, and hip arthroplasty rates.
Animal studies generally showed dose-dependent bone formation and tendon-junction neovascularization (though one cited study suggested delayed bone healing). Clinically, the review quotes success ranges of 34%–88% in plantar fasciitis, 68%–91% in lateral epicondylitis, 78%–91% in calcific shoulder tendinitis, 73.5%–87.5% in patellar/Achilles tendinopathies, and 50%–85% union in non-unions, plus lower 8-to-9-year hip arthroplasty rates after ESWT than core decompression (24% vs 64%, P = 0.002) in one cohort — while also acknowledging blinded trials and a 9-trial Cochrane review in plantar fasciitis and lateral epicondylitis where active ESWT performed similarly to placebo.
No adverse-event rates are tabulated; local reddening, ecchymosis, mild hematoma, and migraine are listed under plantar fasciitis, rabbit tendon damage risk is noted at 0.60 mJ/mm², and complications are repeatedly called 'low and negligible'. The author discloses serving on the scientific advisory committee of Sanuwave (Alpharetta, GA), maker of the OssaTron device, while reporting no manuscript fees.
Unsystematic narrative selection heavily weighted toward the author's own uncontrolled or 6-control cohorts; Sanuwave advisory role; abstract-versus-text mismatch on success rates (65%–91% vs 34%–88% in plantar fasciitis and 50%–85% in non-unions); omission of radial devices from the physics section despite citing radial clinical studies; unpublished personal claims about local anaesthesia; sweeping surgical-replacement conclusions; and two unread 2013 published comments (PMIDs 23895659, 23991453).
Useful only as historical background on early focused shockwave concepts, animal angiogenesis/osteogenesis hypotheses, and 2000–2005 US FDA device approvals. Because uncontrolled series and sham-controlled trials are blended without critical appraisal, it cannot guide clinical decision-making for individual conditions.
Condition-specific systematic reviews and adequately powered, blinded randomized trials that separate focused from radial devices, verify acoustic energy units, account for paired limbs/hips, report confidence intervals, and monitor adverse events systematically.
Read the original article.
Extracorporeal shockwave therapy in musculoskeletal disorders.
Wang CJ. · J Orthop Surg Res · 2012
Open the publication record or available full-text files.
Checked against the full paper · Clinical review: Dr. Marcus Yu Bin Pai, · How we review →